Quick Details
Place of Origin:
Shanghai, China
Brand Name:
Sunlight
Model Number:
SL904P
TDS:
0~10ppt
DO:
0.00~20.00mg/L
Conductivity:
0~20.00, 200.0, 2000uS/cm, 20.00, 200.0mS/cm
Salt:
0.0~200.0%
Temperature:
0-60 Degree
ATC:
Yes
Datalogger:
Yes
USB Communication:
Yes
Supply Ability
Supply Ability:
3000 Piece/Pieces per Month
Packaging & Delivery
Packaging Details
Carrying Case
Port
Shanghai
Lead Time :
3-10 working days after we receive down payment

Handheld Multi-Parameter Water Quality Meter,Conductivity/TDS/Salinity/DO Meter,Datalogger,

USB Communication,SL904P

 

Description:

High accuracy multiparameter water quality meter includes 5 measurement modes, the meter is suitable for measuring the Conductivity, TDS, Salinity, Resistivity and Dissolved Oxygen of the liquids.

 

Conductivity/TDS/Salinity/Resistivity Mode:
• High accuracy portable multiparameter water quality meter is equipped with a large backlit
LCD display.
• 1 to 5 points push-button calibration with automatic calibration solution recognition.
• Selectable cell constants (0.1/1/10), normalization temperatures (20/25°C), TDS conversion
factors, seawater and practical salinity measurement modes.
• Linear or pure water compensation improves measurement accuracy of meter.

 

Dissolved Oxygen Mode:
• 1 or 2 points calibration using air-saturated water and zero oxygen solution.
• Manual salinity and barometric pressure compensation improve the measurement accuracy.

 

Other Features:

• Automatic Temperature Compensation provides accurate reading over the entire range.
• Calibration Due Reminder prompts user to calibrate the meter regularly.
• Stability indicator automatically shows current measurement status.
• Auto-Hold function freezes stable measured value for easy viewing and recording.
• Manual temperature calibration corrects temperature deviation.
• Selectable temperature unit (°C or °F) meets different application requirements.
• Help message as a operational guide that helps you quickly begin using the meter.
• System menu allows setting the 13 parameters, including the calibration points, resolutions,
stability conditions, measurement units, etc.

• Reset feature automatically resumes all parameters back to factory default options.
• Expanded memory stores and recalls up to 500 readings.
• Built-in real-time clock stamps stored data to meet GLP standard.
• Stored data can be transferred into computer by USB communication interface.
• Multi-mode power scheme (battery, power adapter, computer's USB port) ensures that use
the meter smoothly.

 

Specifications:

   Model

SL904P

   Conductivity

   Range

0~20.00, 200.0, 2000µS/cm, 20.00, 200.0mS/cm

   Accuracy

±0.5% F.S

   Calibration Points

1 to 5 points

   Calibration Solutions

10µS/cm, 84µS/cm,1413µS/cm,12.88mS/cm,111.8mS/cm

   Temperature Compensation

0~100°C, 32~212°F, Manual or Automatic

   Temperature Coefficient

0.0~10.0%/°C

   Compensation Modes

Linear or Pure Water

   Cell Constant

K=0.1, 1, 10

   Normalization Temperature

20 or 25°C

   TDS

   Range

0~10ppt (Max. 20ppt, depending on factor setting)

   Accuracy

±1% F.S

   TDS Factor

0.1~1.0 (Default 0.5)

   Salinity

   Range

0~10ppt (Max. 80ppt)

   Accuracy

±1% F.S

   Measurement Modes

Seawater or Practical Salinity

   Resistivity

   Range

0~100MΩ

   Accuracy

±1% F.S

   Resolution

0.01, 0.1, 1

   Dissolved Oxygen

   Range

0.00~20.00mg/L (or ppm)

   Accuracy

±0.2mg/L

   Resolution

0.01mg/L

   Calibration Points

1 or 2 points

   Temperature Compensation

0~50°C, 32~122°F

   Pressure Correction

60.0~112.5kPa, 450~850mmHg

   Salinity Correction

0~50g/L

   % Saturation of Oxygen

   Range

0.0~200.0%

   Accuracy

±2.0%

   Resolution

0.1%

   Temperature

   Range

0~105°C, 32~221°F

   Accuracy

±0.5°C, ±0.9°F

   Resolution

0.1°C

   Calibration Points

1 point

   Calibration Range

Measured value ±10°C

   Others

   Hold Function

Manual or Automatic

   Stability Conditions

Low or High

   Calibration Due

0 to 31 days

   Power Off

Manual or Automatic (10, 20, 30 minutes)

   Reset Function

Yes

   Memory

Stores up to 500 data sets

   Output

USB Communication Interface

   Connector

6-pin Mini Plug

   Power Requirements

3×1.5V "AA" Batteries

   Battery Life

Approximately 150 hours (Turn off the backlit)

   Dimensions

170(L)×85(W)×30(H)mm

   Weight

300g

 

Order Information:

   Order Code

Accessories

   SL904P-CN

CON-1 Conductivity Electrode
 

DO100 Dissolved Oxygen Probe (Made in US)

TP-10K Temperature Probe

Calibration Solutions (84µS/cm, 1413µS/cm, 12.88mS/cm)

DO Filling Solution

DO Cap Membrane

AA Batteries

Carrying Case

   SL904P-UK 

K10 Conductivity Electrode (Made in UK)

DO100 Dissolved Oxygen Probe (Made in US)

TP-10K Temperature Probe

Calibration Solutions (84µS/cm, 1413µS/cm, 12.88mS/cm)

DO Filling Solution

DO Cap Membrane

AA Batteries

Carrying Case

 

Optional Accessories:

Order Code

Description

CON-0.1 Conductivity Electrode

For the pure water

CON-1 Conductivity Electrode

For the general liquids

CON-10 Conductivity Electrode

For the high conductivity liquids

DO100 Dissolved Oxygen Probe

Polarographic sensor

Conductivity Calibration Solutions

84µS, 1413µS, 12.88mS, 111.8mS

DO Filling Solution

500mL

DO Cap Membrane

Special-purpose spare

TP-10K Temperature Probe

Special-purpose sensor

 

Package:

   Order Code

Dimensions

Weight

   SL904P-CN

475(L)×155(W)×350(H)mm

2kg

   SL904P-UK

475(L)×155(W)×350(H)mm

2kg

 

Picture Show: